• Calibration
General
The CND/AR is factory calibrated. Once calibrated it will operate with
any input or output format without recalibration.
Allow the instrument 15 minutes of powered operation (to reach a
stable temperature) before calibration and allow a few seconds for
signals to stabilise before accepting their value.
Equipment requirements
• Suitable precision 50 Ω, 200 Ω, 1 kΩ, 5 kΩ resistors (±0.05 % or
better accuracy)
• An Rtd simulator with ±0.05 % or better accuracy
• An accurate digital multimeter (accurate to 0.05 mV and ±0.1 μA)
Terminal connections for input calibration
Calibration Stage Signal type Terminal
Calibrating resistance
2
3
Pt100 RTD simulator
A 4
B 5
Bsense 6
Analogue Current Output
mA output + 14
mA output − 15
Analogue Voltage Output
V output + 14
V output − 16